Suppose a US distributor places a first commercial order for sterile single-use surgical consumables on a Delivered Duty Paid (DDP) quote to a regional warehouse. Two weeks before dispatch, the seller’s customs broker sends a draft entry sheet asking for a US entity name, address and FEI to put in the FDA Device Initial Importer box. Purchasing assumed DDP meant the supplier handled every US import formality. Warehouse operations assumed the third-party logistics (3PL) address on the bill of lading solved it. Neither assumption holds up if the draft entry goes to U.S. Customs and Border Protection (CBP) and the Food and Drug Administration (FDA) with the wrong party.16

This guide gives US distributors, foreign medical-device suppliers and import operations teams a pre-dispatch reconciliation method for ordinary commercial shipments. It maps the purchase order, statutory customs eligibility under 19 U.S.C. § 1484, FDA establishment definitions under 21 CFR Part 807, and the March 10, 2026 edition of FDA’s Automated Commercial Environment (ACE) quick reference guide so the buyer and seller can resolve entity mismatches before cargo leaves the factory.235

Contents: 1. What DDP requires · 2. CBP importer vs. FDA initial importer · 3. FD1, DII and DP in ACE · 4. Pre-dispatch worksheet · 5. Three worked DDP cases · 6. Postmarket duties · Where VEMERIX fits · FAQ · Sources and scope

What does a DDP quote require the seller to arrange?

Incoterms 2020 DDP gives the seller responsibility for export, transit and import clearance where applicable, including applicable duties and taxes. Delivery and risk transfer take place at the named destination, with the cleared goods on the arriving vehicle and ready for the buyer to unload. Specify the receiving point on the purchase order so the parties can match it to the shipping instructions.1

Two details in the ICC text directly affect medical-device buyers reviewing a DDP quote before issuing a purchase order:

  • Buyer assistance under Article B7. If the seller requests help obtaining documents or information for clearance, the buyer assists where applicable at the seller’s risk and cost. Providing a product code or listing number to the broker gives the filer supporting data. Identify the actual importer of record separately.12
  • When the buyer needs to clear the goods. ICC’s DDP Explanatory Note 7 suggests considering Delivered at Place (DAP) or Delivered at Place Unloaded (DPU) if the seller cannot obtain import clearance and the buyer is to handle it. Confirm which party will clear the goods and how the new terms affect the quote before releasing the order.1

Incoterms 2020 allocates delivery risk, carriage and clearance costs between buyer and seller. For US entry, verify statutory eligibility and FDA establishment roles from the parties’ actual activities and records. If you are also comparing EXW, FOB and DDP pricing, use the landed-cost quote worksheet for the price comparison.1234

Which entity qualifies as the CBP importer of record versus the FDA device initial importer?

Customs eligibility and FDA device importer status are governed by different titles of federal law. One party can hold both roles on a single shipment, or two different parties can hold them, but each designation must pass its own statutory or regulatory test independently.23

CBP Importer of Record under 19 U.S.C. § 1484

For ordinary US commercial entry, 19 U.S.C. § 1484(a)(1) requires one of the parties qualifying as “importer of record” under paragraph (a)(2)(B)—either in person or through an agent authorized in writing—to make and complete entry using reasonable care.2 Under § 1484(a)(2)(B), required entry documentation or electronic data must be filed or transmitted either by the owner or purchaser of the merchandise or, when appropriately designated by the owner, purchaser or consignee, a person holding a valid customs broker licence under 19 U.S.C. § 1641.2 The statute states expressly that the importer of record must be one of the parties eligible to file under that section.2

For an electronic entry, the importer of record or its agent certifies each transmission. At least one of them must be a US resident for service of process.2 A nonresident corporation entering goods for consumption also needs an authorized resident agent in the applicable state and a bond with a resident corporate surety under 19 CFR § 141.18. Ask the broker to document the eligibility basis, bond and any power of attorney that applies. A freight forwarder or warehouse that merely handles the goods has no owner or purchaser eligibility on that basis.216

FDA Device Initial Importer under 21 CFR 807.3(g) and 807.20(a)(5)

Under 21 CFR § 807.3(g), the FDA initial importer furthers marketing from the foreign manufacturer toward final delivery or sale to the device’s user. This role covers devices whose container, wrapper and labeling the importer leaves unchanged. Read that provision with § 807.3(x), which defines a Part 807 importer as a US company or individual that owns, receives or takes consignment of the imported device. The definition generally excludes the end consumer or patient, with an exception for direct shipments from the foreign establishment.3

Appendix A of FDA’s March 10, 2026 Industry Quick Reference Guide also specifies a physical US address staffed by people responsible for the compliance of the imported devices. Check who handles that work at the address the broker will use.5

Under 21 CFR § 807.20(a)(5) and FDA’s establishment registration table (page current as of November 19, 2026), an initial importer must register annually with FDA and pay the annual establishment registration fee.410 However, § 807.20(a)(5) gives initial importers a narrow way to satisfy their listing duty: for any device for which they did not initiate or develop specifications and did not repackage or relabel, they fulfill their listing obligation by submitting the name and address of the manufacturer—summarized on FDA’s registration table as “Identify manufacturers per 807.20(a)(5)”—while remaining prepared to submit the proprietary name (if any) and common or usual name of each imported device upon FDA request.410

Adjacent roles that do not substitute for DII

Before accepting any entity name on a draft customs entry, verify that a nearby regulatory or logistics label has not been misread as Device Initial Importer:

  • Foreign establishment’s U.S. Agent (21 CFR § 807.3(s)). A registrable foreign establishment designates a U.S. agent with a place of business or residence in the United States and an individual who is physically present. The agent handles FDA communications. For DII, test separately whether the agent also meets the marketing and unchanged-device conditions of § 807.3(g).37
  • Person who imports or offers for import (21 CFR 807.3(y)). Section 807.3(y) defines this as an agent, broker or other entity (other than a carrier) used by the foreign establishment to facilitate the import of its device into the United States.3 FDA’s domestic registration table explicitly marks “NO” for registration, listing and fee for a broker or facilitator that does not further marketing under § 807.3(g) and § 807.20(a)(5).10
  • Wholesale distributor (21 CFR 807.3(t) & 807.20(c)). A wholesale distributor is a person other than the manufacturer or initial importer who distributes a device from the original place of manufacture to the person making final delivery or sale.3 If it does not manufacture, repackage, process or relabel, § 807.20(c) exempts it from Part 807 registration and listing.4 A downstream US wholesaler buying from a registered initial importer does not automatically become the DII on the import entry.34
  • Repackager or relabeler (21 CFR §§ 807.3(d)(1) and 807.20(a)(3)). Repackaging or changing label content for distribution can require registration and listing as a repackager or relabeler. FDA’s establishment descriptions exclude placing finished packages in shipping cartons for transport and, for relabelers, adding an establishment name while otherwise leaving the original label unchanged. Have regulatory staff assess the exact change for the shipment before relying on initial-importer status.3410
  • Delivered to Party (DP) / passive 3PL warehouse. In FDA’s ACE glossary, DP simply identifies the US facility that physically receives the goods after arrival.5 A fulfillment warehouse receiving pallets does not satisfy § 807.3(g) marketing furtherance or the staffed compliance responsibility required for DII.35
Role / LabelControlling AuthorityQualifying TestUS Address & Registration DutyWhat It Does Not Prove
DDP SellerIncoterms 2020 DDP (sale contract)1Seller agrees to deliver goods cleared for import at named destination point.1No automatic US entity or Part 807 initial-importer status.13Does not prove CBP IOR eligibility or FDA product admissibility.27
CBP Importer of Record / FDA Importer (FD1)19 U.S.C. § 1484; 21 CFR 1.72(b); ACE Guide258Owner, purchaser or properly designated licensed broker. For electronic entry, the IOR or agent certifies the data, with at least one US resident for service of process.2FDA’s entity table requires an individual contact name, telephone and email. Part 807 initial-importer registration depends on the activity.3458Does not substitute for DII when another US party furthers marketing.35
Device Initial Importer (DII)21 CFR 807.3(g), (x) & 807.20(a)(5); ACE Appendix A345US owner, consignee or recipient furthering marketing from foreign manufacturer toward final delivery/sale without repackaging or relabeling.3Staffed physical US address; must register annually, pay fee and identify manufacturers (plus MDR, Part 806 & complaint duties).457Does not prove the device is cleared/approved, nor authorize US relabeling.311
Foreign Establishment U.S. Agent21 CFR 807.3(s) & FDA Import Guidance37US-resident or US-place-of-business agent designated for FDA communications on behalf of a foreign establishment.3Physical US presence (no mailboxes or answering services); named in foreign establishment’s registration.37Does not make the agent a CBP IOR or an FDA DII.3
Import Facilitator (Broker / Agent)21 CFR 807.3(y) & Registration Table310Agent, broker or entity (other than carrier) facilitating import for foreign establishment without furthering marketing.310No Part 807 registration, listing or fee if activity stays within § 807.3(y).10Cannot be copied into DII solely because it transmits the entry.610
Delivered to Party (DP)FDA ACE Quick Reference Guide, p. 25 & p. 355Physical US site where goods are delivered after arrival (e.g., buyer DC or 3PL warehouse).5Entity name and US address mandatory in ACE; no registration duty from storage alone.5Physical receipt does not qualify a 3PL or warehouse as DII.35
Table 1. Commercial, customs and FDA device roles on a US DDP import

Source: ICC Incoterms 2020 [1], 19 U.S.C. § 1484 [2], 21 CFR Part 807 [3–4], and FDA’s ACE and registration guidance [5, 7, 10].

Where do the customs importer, device initial importer and warehouse appear in FDA entry data?

When a customs broker files a medical-device entry in CBP’s Automated Commercial Environment (ACE), FDA reviews the Partner Government Agency (PGA) message set electronically.711 According to CBP bulletin CSMS #68020465 (March 12, 2026), the changes in FDA ACE Supplemental Guide Version 2.6 entered production on March 10, 2026, accompanied by FDA’s updated March 10, 2026 Industry Quick Reference Guide to the FDA ACE Supplemental Guide.512 Your broker should always confirm the current production CATAIR specification at filing.

In the guide’s medical-device entity table on p. 25, FDA shows five roles with mandatory entity names and addresses. The filer’s or broker’s point of contact is an optional sixth role.5

  • MF (Manufacturer) and DEQ (Shipper). Entity name and address are mandatory; individual contact fields are N/A; DUNS or FEI is optional, though encouraged.5 FDA emphasizes on its Common Entry Submission Errors page (updated March 5, 2026) that FDA’s manufacturer means the site-specific location where the device is made or produced, which can differ from a CBP manufacturer declaration.6
  • FD1 (FDA Importer). This is the FDA message-set field for the importer of record. The guide requires the entity name and address as well as an individual name, telephone number and email. The regulation at 21 CFR § 1.72(b) specifies the telephone and email. An entity ID (DUNS or FEI) is encouraged and optional.58
  • DII (Device Initial Importer). Entity name and address are mandatory; individual name, tel# and email are N/A; DUNS or FEI is optional, though encouraged.5
  • DP (Delivered to Party). Entity name and address are mandatory (must be a US location per FDA’s common-errors guidance); individual contact fields are N/A; DUNS or FEI is optional, though encouraged.56
  • PK (Filer’s/Broker’s Point of Contact). Both entity name/address and individual name, tel# and email are optional, though encouraged, to speed up FDA review questions.5

Two subtle points in that March 2026 entity table trip up procurement teams. First, FDA notes under the table that “Optional, but encouraged” entity IDs (DUNS or FEI) assist database matching and entry review, and that CDRH prefers FEIs for devices.5 Treat that as a matching aid, not a rule that makes FEI transmission mandatory in the entity block—and conversely, never assume an optional entity ID field means the initial importer can skip FDA registration under 21 CFR § 807.20(a)(5).45

The broker also needs a separate product-line check under 21 CFR §§ 1.72 and 1.76. Table 2 uses the standard foreign-manufactured finished-device scenario from p. 24 of FDA’s guide, with Intended Use Code (IUC) 081.001. The guide also lists other scenarios under that code. Have the broker confirm the scenario for the actual shipment.589

ACE Element / CodeMarch 2026 Status (IUC 081.001)Controlling BasisPre-Dispatch Buyer / Broker Verification
FDA Country of ProductionMandatory21 CFR § 1.72(a)(1); FDA guide p. 2658Confirm the country of the last manufacturing or processing that had more than a minor effect on the article.
Complete FDA Product Code & Invoice DescriptionMandatory21 CFR § 1.72(a)(2); FDA guide p. 23; FDA entry errors568For the worked scenario, confirm the DEV program and NED processing code. Check that the complete FDA product code is consistent with the invoice description. Split FDA lines when a required data element differs, such as manufacturer or product code.
Full Intended Use Code (081.001)Mandatory21 CFR § 1.72(a)(3); FDA guide p. 2458Confirm that the entry is a standard import of a foreign-manufactured finished device. Check the applicable scenario separately for samples, convenience kits and Import for Export.
DEV (Foreign Manufacturer Registration) & DFE (Foreign Exporter Registration)Mandatory21 CFR § 1.76(a); FDA guide p. 24; FDA code definitions591314Have the broker match each registration to the relevant establishment. FDA allows the applicable owner/operator number while a registration number is being assigned.
LST (Device Listing Number)Mandatory21 CFR § 1.76(a); FDA guide p. 24; FDA device entry errors5914Request the LST number from the official correspondent of the firm that lists the device. FDA’s public database shows listings and does not display the proprietary LST number.
PM# (Premarket Number)Conditional (mandatory if required for the device)21 CFR § 1.76(c); FDA guide pp. 23–24591114Match the applicable premarket number to the device and listing. FDA distinguishes a 510(k) clearance, PMA approval, De Novo authorization, completed PDP and HDE.
IRC / LWC / DIIRC & LWC Conditional; DI Optional21 CFR § 1.76(e)–(f); FDA guide p. 24; FDA code definitions5913IRC is the impact-resistance affirmation for applicable eyeglass or sunglass lenses. LWC applies to covered electrode lead wires or patient cables. DI is the device identifier portion of the UDI and is optional on this entry line.
FD1, DII, DP Entity Records & Anticipated ArrivalMandatory entity names/addresses & arrival date/time21 CFR § 1.72(b); FDA guide pp. 25–26 and 35; FDA entry guidance56817Confirm the FD1 contact, DII street address and physical DP receiving site. FDA says CBP transmits the entry for FDA review once the current date is within five days of the declared arrival date.
Table 2. Standard finished-device ACE entry elements (Intended Use Code 081.001, March 10, 2026 FDA quick reference)

Source: 21 CFR §§ 1.72 and 1.76 [8–9], FDA Industry Quick Reference Guide, pp. 23–26 [5], FDA ACE Affirmations of Compliance [13] and FDA entry error guidance [6, 14].

FDA uses PREDICT for electronic screening and can compare submitted entity and product identifiers against registration, listing and premarket records. An incorrect DII or a line that combines different manufacturers can cause additional review. An incorrect DP can send FDA communications to the wrong party.611

Which documents should agree before the broker submits the entry?

Do not wait until ocean or air freight is in transit to discover that the seller’s broker copied a warehouse into DII or assumed the buyer would act as FD1 under a DDP price. Run this six-row pre-dispatch reconciliation across the commercial quote, the importer eligibility files and the broker’s draft ACE line before authorizing factory release.

Record / Entry BlockWhat Must Be Verified Before DispatchSource Evidence to RequestAccountable OwnerHold Trigger / Stop Condition
1. DDP quote and purchase orderIncoterms 2020 rule, exact named US delivery point, seller import-clearance plan, and who supplies entry documents or answers FDA/CBP queries (B7/A9).1Signed quote and PO specifying DDP named place, clearance responsibility and document-assistance terms.Seller commercial lead & buyer purchasing leadHold if clearance is unassigned or the quote conflicts with the broker-confirmed plan.
2. CBP Importer of Record & FDA FD1Record the eligibility basis under 19 U.S.C. § 1484, the bond and authorization as applicable, the resident party for service of process and the individual contact details required in FD1. If a nonresident corporation will enter goods for consumption, verify § 141.18 as well.25816Written broker confirmation of IOR statutory basis, bond/POA status and redacted FD1 draft entry block.Seller logistics lead & licensed customs brokerHold if the IOR’s legal basis is unverified or the broker’s authority only covers filing on behalf of a different IOR.
3. FDA Device Initial Importer (DII)US entity (owner, consignee or recipient) actually furthering marketing from the foreign manufacturer toward final sale/delivery without changing container, wrapper or labeling; staffed US physical address; active annual registration and manufacturer identification.345Written commercial-role confirmation from the US initial importer, current FDA registration proof, manufacturer link and broker draft DII row.US initial importer & regulatory compliance leadHold if the only evidence of eligibility is a service-provider label, receiving address or FEI. If the buyer plans to repackage or relabel, resolve the actual regulatory role before using this DII assignment.
4. Delivered to Party (DP)Physical US receiving facility and street address match shipping instructions, booking and ACE draft.56Delivery instructions, air/ocean booking and broker draft DP block.Warehouse / receiving lead & logistics coordinatorCorrect any destination mismatch before filing and recheck whether a changed receiver alters DII or FD1.56
5. Product entry data (081.001)Match the invoice description to the FDA product code and the actual manufacturer site, then confirm the 081.001 scenario, DEV, DFE, LST and the applicable PM#.56913Manufacturer technical/regulatory summary, private-share LST confirmation, public registration/510(k) check and line-level broker draft.Foreign manufacturer QA/RA & licensed customs brokerHold if any model lacks verified registration/listing or required PM#, or if multiple product codes or manufacturers are lumped on one line.6911
6. Postmarket handoffNamed US regulatory contact for complaint intake and forwarding to manufacturer, 21 CFR 803 MDR reporting, 21 CFR 806 corrections/removals and 21 CFR 821 tracking where applicable.7Signed distributor or quality-role record assigning complaint escalation, MDR event files and recall communication contacts.US initial importer & manufacturer quality leadsConfirm who receives complaints and evaluates reportability before the first shipment.
Table 3. Pre-dispatch DDP quote-to-entry reconciliation worksheet for ordinary US medical-device imports

Source: VEMERIX worksheet based on ICC Incoterms 2020 [1], US customs and FDA rules [2–4, 8–9, 15–16], and FDA’s entry guidance [5–7, 11, 13–14, 17].

When should the buyer hold dispatch?

Use the same checks in each example. Start with the contract, then work through the customs importer, the company marketing the device in the United States and the delivery site. Check the product line before signing off on dispatch.

These examples use fictional shipments of finished sterile single-use devices. Verify the device’s actual identity, FDA status and import scenario for a live order.

Case A: Distinct entities for FD1, DII and DP (Valid separation)

A foreign manufacturer sells finished, 510(k)-cleared single-use wound-closure devices to a US specialty distributor, DDP to a named Newark warehouse under Incoterms 2020. In this example, the broker confirms that the seller has a qualifying ownership interest, the required bond and an authorized resident agent in the applicable state. The seller is entered in FD1.12516 The US distributor buys and markets the devices in their original packaging, has a staffed US compliance address, is registered as the initial importer and identifies the manufacturer. The distributor is entered in DII. The contracted warehouse that takes physical delivery goes in DP.345

Dispatch decision. The buyer can sign off on the role mapping once the broker verifies the product line, including the model’s applicable PM#, and the distributor confirms who handles complaints. CBP and FDA make their own decisions at entry.5711

Case B: One US distributor in both FD1 and DII (Valid overlap)

The supplier initially quotes DDP, but cannot arrange import clearance. The parties agree to DAP at the named US destination and revise the price to reflect the buyer handling clearance. The US distributor, as purchaser, is the broker-verified importer of record and appears in FD1.125 The distributor also markets the unchanged device, maintains a staffed US compliance address and registers as initial importer, so it appears in DII. Its regional fulfillment center appears in DP.345

Dispatch decision. Accept the matching names on FD1 and DII once the broker and regulatory leads have signed off on each role and the product line.235

Case C: Warehouse copied to DII plus planned US relabeling (Hold for correction)

A US private-label buyer orders finished sterile single-use devices on a DDP quote. The broker’s draft copies a passive fulfillment warehouse from a previous entry into both DP and DII. The buyer also plans to change the brand and UDI on the actual device label at its own facility before selling to clinics.

Dispatch decision. Hold and escalate both discrepancies. The warehouse goes in DP if it receives the goods. Ask the US regulatory lead to identify the actual marketing party and assess the proposed label and UDI changes. If the buyer will act as a relabeler, confirm the registration and listing for that activity, the applicable device and labeling requirements and the correct FDA entry treatment with the broker before approving shipment. If the supplier instead delivers the final labeled device, reassess DII based on who markets that configuration and verify its model-specific entry data.3451011

Worked CaseContract TermFD1 (CBP IOR)DII (FDA Initial Importer)DP (Delivery Site)Dispatch Decision & Owner
Case A: Distinct entitiesDDP US destination; seller handles clearanceForeign seller with broker-confirmed eligibility, bond and resident agentRegistered US distributor marketing unchanged wound-closure devices from its staffed US officeContracted 3PL warehouseProceed after confirming device-line data and the complaint contact.
Case B: Same-entity overlapDDP quote revised to DAP for buyer clearanceUS distributor (purchaser under § 1484(a)(2)(B))Same US distributor (furthers marketing without relabeling under § 807.3(g))Separate fulfillment centerProceed once both roles and the product line are verified.
Case C: Warehouse DII + relabelingDDP US destinationUnresolved or misaligned with commercial plan3PL warehouse copied from stale template while buyer plans US relabeling3PL warehouseHold while regulatory staff resolve the DII assignment, relabeling and device evidence for the broker.
Table 4. How three US DDP shipment patterns resolve under pre-dispatch review

Source: Hypothetical scenarios based on ICC Incoterms 2020 [1], US customs and FDA rules [2–4, 16], and the FDA ACE quick reference guide [5].

Postmarket duties that stay with the initial importer after customs release

Assign postmarket responsibilities to the company acting as the initial importer before shipment. For the distributor, the following contacts and records belong in the quality handoff.715

  • Annual establishment registration (21 CFR Part 807). Renew registration annually (including the annual registration fee) and maintain accurate manufacturer identifications.4710
  • Medical Device Reporting (21 CFR § 803.40). The importer reports qualifying deaths or serious injuries to FDA and sends a copy to the manufacturer. It reports a qualifying malfunction to the manufacturer when a recurrence would be likely to cause or contribute to death or serious injury. Apply the applicable reporting procedures and maintain MDR event files.715
  • Complaint forwarding to the manufacturer. All product complaints—including both MDR-reportable and non-MDR events—must be forwarded by the importer to the manufacturer.7
  • Corrections, removals and tracking (21 CFR Parts 806 & 821). Importers are subject to Corrections and Removals reporting under 21 CFR Part 806 and Medical Device Tracking under 21 CFR Part 821 where applicable to the device.7

If a shipment is detained after filing, follow the notice and confirm whether FDA authorization is needed for any proposed relabeling or reconditioning. See our guide to FDA detention and reconditioning. For a different transaction involving processing and re-export, see the guide to the US Import for Export route.711

Where VEMERIX fits—and where due diligence still begins

VEMERIX is the international brand of Weihai Medison Medical Equipment Co., Ltd., positioned as a Minimally Invasive Surgery Total Solution Platform serving urology, vascular surgery and perioperative care. When an international distributor or OEM buyer evaluates a sterile single-use product line, VEMERIX supports pre-order due diligence by supplying manufacturer specifications, quality system records and available regulatory documentation for the specific SKU and target market.

Before approving a US purchase order, have your customs broker verify FD1 and the entry data, and have your US regulatory lead confirm the initial importer’s activities and the exact device’s US market status.

To review available manufacturer quality records or request a model-specific technical pack for a prospective distribution discussion, visit VEMERIX quality and regulatory documentation or submit your target market, product family and intended distribution flow through the VEMERIX contact page. For broader supplier-file verification before placing a first commercial order, you can also review our sample-to-order supplier evidence workpaper.

Frequently asked questions

Can the same US distributor be the customs importer and the device initial importer?

Yes. The distributor can appear in FD1 as an eligible US purchaser and in DII when it also markets the unchanged device from a staffed US compliance site and meets its registration obligations. Keep the evidence for both roles in the shipment file.2345

Can the broker, US agent or delivery warehouse be entered as the initial importer?

Not unless that entity actually satisfies 21 CFR § 807.3(g) by furthering the marketing of the device from the foreign manufacturer toward final delivery or sale without changing the container, wrapper or labeling, and maintains the staffed US compliance address described in FDA’s ACE glossary.35 A customs broker acting under § 807.3(y), a foreign establishment’s U.S. agent acting under § 807.3(s) or a passive warehouse receiving freight in DP does not automatically qualify.3510

Does the initial importer submit a separate device listing?

Under 21 CFR § 807.20(a)(5), an initial importer must register and pay the annual fee, but may fulfill its listing obligation for devices for which it did not initiate or develop specifications and did not repackage or relabel by submitting the manufacturer’s name and address (and remaining ready to provide the proprietary and common/usual names on FDA request).410 On the ACE product line, however, the filer still transmits the applicable DEV, DFE and LST numbers for the device.5911

What should change if the US party plans to relabel or repackage?

Pause the shipment review and document the proposed change. A company that repackages or relabels the device is subject to the registration and listing requirements for that activity. Confirm how the changed labeling, any UDI changes and the resulting entry scenario affect the exact device before the broker revises the draft.34510

What if the DDP seller cannot arrange an eligible importer of record?

ICC’s Incoterms 2020 Explanatory Note 7 for DDP explicitly advises considering DAP or DPU when the seller cannot obtain import clearance and the parties intend the buyer to handle import clearance in the destination country.1 Adjust the contract terms and pricing allocation before dispatch—and still confirm separately who qualifies as FD1 and DII.123

Source scope and when to recheck

Sources were checked November 24, 2026. The field crosswalk uses Section 7 and Appendix A of FDA’s March 10, 2026 Industry Quick Reference Guide and the October 14, 2026 FDA code definitions. CBP confirmed that FDA Supplemental Guide v2.6 entered production on March 10, 2026.51213

The worked example covers ordinary US commercial entry of finished, non-radiation-emitting CDRH devices. The examples use hypothetical transactions. Have the broker confirm the live CATAIR specification and the scenario for the actual line, especially for convenience kits, components, investigational devices, refurbished or reprocessed devices, or Import for Export. Recheck the worksheet whenever the Incoterm, importer, warehouse, labeling, manufacturer site or FDA guide changes.512

Sources

  1. International Chamber of Commerce (ICC), Incoterms 2020 Rules for Any Mode of Transport — DDP (Delivered Duty Paid), Explanatory Notes 1, 3, 6–7 and Articles A2, A7, A9 and B7 (printed pp. 83–89; accessed November 24, 2026).
  2. U.S. Government Publishing Office, 19 U.S.C. § 1484 — Entry of merchandise (United States Code, 2024 Edition; accessed November 24, 2026) — subsections (a)(1), (a)(2)(B) and (d)(1) on importer-of-record eligibility, reasonable care and U.S. resident certification for service of process.
  3. Office of the Federal Register and Government Publishing Office, 21 CFR § 807.3 — Definitions (Electronic Code of Federal Regulations, Title 21 current as of November 20, 2026; accessed November 24, 2026) — paragraphs (d)(1), (g), (s), (t), (x) and (y).
  4. Office of the Federal Register and Government Publishing Office, 21 CFR § 807.20 — Who must register and submit a device list? (Electronic Code of Federal Regulations, Title 21 current as of November 20, 2026; accessed November 24, 2026) — paragraphs (a)(3), (a)(5) and (c).
  5. U.S. Food and Drug Administration (FDA), Industry Quick Reference Guide to the FDA ACE Supplemental Guide (March 10, 2026 edition; accessed November 24, 2026) — Section 7 Medical Devices (pp. 23–26, including scenario 081.001 on p. 24 and the entity table on p. 25) and Appendix A Entity Glossary (p. 35).
  6. U.S. Food and Drug Administration (FDA), Common Entry Submission Errors (page content current as of March 5, 2026; accessed November 24, 2026) — entries on incorrect device initial importer, consignee, manufacturer, product code, intended use code, line grouping and arrival date.
  7. U.S. Food and Drug Administration (FDA), Importing Medical Devices and Radiation-Emitting Electronic Products into the U.S. (page content current as of January 30, 2026; accessed November 24, 2026) — Foreign Manufacturers, Initial Importers of Medical Devices and Import Process.
  8. Office of the Federal Register and Government Publishing Office, 21 CFR § 1.72 — Data elements that must be submitted in ACE for articles regulated by FDA (Electronic Code of Federal Regulations, Title 21 current as of November 20, 2026; accessed November 24, 2026) — paragraphs (a)(1)–(3) and (b).
  9. Office of the Federal Register and Government Publishing Office, 21 CFR § 1.76 — Medical devices (Electronic Code of Federal Regulations, Title 21 current as of November 20, 2026; accessed November 24, 2026) — opening paragraph and paragraphs (a), (c) and (e)–(g).
  10. U.S. Food and Drug Administration (FDA), Medical Device Establishment Registration: Who Must Register, List and Pay the Fee (page content current as of November 19, 2026; accessed November 24, 2026) — domestic and foreign establishment tables and role definitions.
  11. U.S. Food and Drug Administration (FDA), Importing Medical Devices (accessed November 24, 2026) — how FDA verifies compliance via PREDICT, registration and listing verification, and premarket submission matching.
  12. U.S. Customs and Border Protection (CBP), CSMS #68020465 — REMINDER: Production Deployment Date for FDA Supplemental Guide v. 2.6 was March 10, 2026 (issued March 12, 2026; accessed November 24, 2026).
  13. U.S. Food and Drug Administration (FDA), FDA ACE Affirmations of Compliance (October 14, 2026 edition) — DEV, DFE, DI, IRC, LST, LWC and PM# definitions and qualifiers.
  14. U.S. Food and Drug Administration (FDA), Medical Device Common Entry Errors — registration, listing and premarket number lookups.
  15. Electronic Code of Federal Regulations, 21 CFR § 803.40 — Medical Device Reporting requirements for importers.
  16. Electronic Code of Federal Regulations, 19 CFR § 141.18 — Entry by nonresident corporation.
  17. U.S. Food and Drug Administration (FDA), Entry Submission — FDA Establishment Identifiers, entry data and arrival timing.

Talk to VEMERIX